Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Would any problems arise if I wired the current input to both the direct input terminal and external sensor input terminal?

0
10 Posted

Would any problems arise if I wired the current input to both the direct input terminal and external sensor input terminal?

0
10

You should not use the instrument with both the direct input and external sensor terminals wired. There would be a risk of electric shock or damage to the instrument since the external sensor input terminal’s Lo side, and the Lo side (+/- terminal side) of the current input terminal for direct input would both be at the same electric potential. When inputting a measurement signal to the direct input terminal, you must remove the shunt connected to the external sensor input terminal, and the current sensor.

Related Questions

What is your question?

*Sadly, we had to bring back ads too. Hopefully more targeted.