Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Why is ICR inherently able to provide better resolution and mass accuracy than conventional mass spectrometers?

0
Posted

Why is ICR inherently able to provide better resolution and mass accuracy than conventional mass spectrometers?

0

The main reason is that a superconducting magnet has a stability of a few parts per billion (ppb) per hour. Quadrupole and TOF instruments each require a voltage of ~1-10 keV, and it is (so far) not possible to stabilize such a voltage to better than ~1 part in 10,000. Ergo, FT-ICR has 10-100X higher resolving power.