Are temperatures of sample and displayed one the same?
• Displayed temperature is the one of right on the pan, that is, of sample especially when spread on the pan evenly. However when placed unevenly or it has hard skin membrane the temperature is of the surface of the sample. MX/MF does not have the sensor right on the pan but calculates the place whose temperature is assumed to be the one of on the pan and puts the sensor there. Temperature calibration is possible by users. Back to top • Why moisture content rate MX/MF measured is less than what FD-620 measured and does the former take shorter time to measure? • With FD-620, temperature displayed is lower than the actual and thus, it heats shorter and value can be lower. • MX/MF is an improvement over FD-620 in the sense of sensing temperature of the pan. That is, displayed temperature is assumed to be the one of on the pan. To get the same result from a sample lessen by 1g and set complete condition at HI. Back to top • What is highly accurate SHS like? • SHS stands for Super Hybrid Se