Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Why is the verification dose experiment performed at a lower SAL than the sterilization dose?

0
Posted

Why is the verification dose experiment performed at a lower SAL than the sterilization dose?

0

In order to test a dose for SAL 10-6, one million products would need to be irradiated and sterility tested. SAL 10-2 (Method 1) or SAL 10-1 (VDmax) is used because only 100 or 10 products, respectively, must be used for the experiment.