Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Why doesn the AP4 provide direct readings of transpiration rate?

rate readings transpiration
0
Posted

Why doesn the AP4 provide direct readings of transpiration rate?

0

We are frequently asked about making measurements of transpiration with the AP4. It is possible to convert from conductance readings to transpiration readings, and the basic equation used to define conductance will do this (see the AP4 manual, page 85): E = cv (wvdc – wvdl) where E = flux density of water vapour, in g m-2 s-1, cv = conductance in velocity units, m s-1, wvdc = water vapour density, at the cup RH and temperature, g m-3. wvdl = water vapour density, at the leaf RH and temperature, g m-3. However, this measurement will only apply to the conditions within the cup itself. It cannot be used as a basis for calculating the transpiration loss from a plant or canopy because the leaf temperature, air temperature, vapour pressure deficit and boundary layer conductance will all be different in the cup from in the field. We don’t include the option to measure transpiration rate in the AP4 so that users are not tempted to make this extrapolation. In fact, although the ability to take