Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Why do I get red error events in WinCal when I calibrate the PNA for fewer ports than are displayed on the PNA?

0
Posted

Why do I get red error events in WinCal when I calibrate the PNA for fewer ports than are displayed on the PNA?

0

WinCal XE has this limitation, which will be fixed in a future release. The error set calculated must be applicable to the displayed data on the PNA. If, for example, you have a quad display with all four 2-port S-parameters on the PNA and perform a 1-port SOL, you will get an error message in WinCal (red event) when the PNA tries to correct S21 (cannot be done with SOL). WinCal will consider the calibration a failure, but S11 on the PNA is indeed calibrated.

Related Questions

What is your question?

*Sadly, we had to bring back ads too. Hopefully more targeted.

Experts123