Why can we reach the high temperature/high humidity condition of 85C and 95% humidity?
A. Some users have found that when they first turn on the chamber, the temperature levels out at 85C and the humidity goes to 95%. After about ten minutes the low stage compressor starts up, and the humidity inside the chamber goes down to about 60%. The steam generator is full on when this happens but they can’t reach the condition that they need. Make sure that the Channel 11 output is set to “Time Proportioning Control” and not “On/Off”. “Time Proportioning Control” prevents the ambient coil from being on all of the time which reduces the chamber humidity, thus allowing the chamber to reach a higher humidity. Channel 11 can be controlled from the Events Screen while running a program as well as during steady state mode.