Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Why can my Refrigerated Multipurpose Centrifuge 5804 R (or 5810 R) maintain 4°C at maximum speed?

0
Posted

Why can my Refrigerated Multipurpose Centrifuge 5804 R (or 5810 R) maintain 4°C at maximum speed?

0

The Multipurpose Centrifuges 5804 R and 5810 R can maintain 4°C at maximum speed with all rotors except the 6 x 85 ml Fixed-angle Rotor (022627207). If 4°C cannot be maintained at maximum speed with rotors other than the 6 x 85 ml Fixed-angle Rotor, please contact Brinkmann Instruments Services. The following specifications apply to the 6 x 85 ml Fixed-angle Rotor only: • Models 5804 R and 5810 R can maintain 4°C with this rotor at 10,500 rpm • Model 5804 R can maintain 6°C at maximum speed • Model 5810 R can maintain 12.