Why can I download a profile to my Partlow 1460/62 controller?
A. After establishing proper communications, the Input code for the Partlow 1460/62 must be entered into Synergy Manager before downloading a profile. This Input code defines the temperature range and engineering units designation needed for compiling the profile. This code is not available from the Partlow 1460/62 through communications and therefore must be provided by the user manually. To enter code, click the ‘Events’ tab on chamber window. Then click the ‘Controller Ranges’ sub tab button. Then click the ‘Enter Input Setup Code’ button. At this point a numeric keypad will appear, just enter proper code and click the ‘Ok’ button. The proper code can be found using the Partlow 1460/62 user manual and verifying the Partlow 1460/62 configuration. Note: Synergy Manager version 2.09.x and later now automatically sets code to ‘7222’. If input code 7222 is not correct, use the procedure above to set the proper code as per users Partlow 1460/62 configuration.