Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Why can I not calibrate the external cell linked to my VR3 or why does it show a different reading to my rebreather cells?

0
Posted

Why can I not calibrate the external cell linked to my VR3 or why does it show a different reading to my rebreather cells?

0

It is likely that the rebreather has not been properly calibrated. To test this, follow the manufacturers’ recommended calibration procedure. For example:- · If you are at sea level set the atmospheric pressure at 1000mb; if not adjust for altitude · Fill the rebreather completely with oxygen with the loop closed · Then evacuate the loop, allowing no air in at any time · Repeat this three times · After the last fill make sure the loop is at ambient pressure by momentarily opening and closing the mouth piece or pulling the counter-lung dump cord · Read the rebreather displays · If they are at 1.0 Bar PO2, the rebreather is correctly calibrated · If not, repeat the calibration process NOTE: you may get a cell stuck message while calibrating with the counterlungs full of oxygen. If so, you may need to add a breath of air to the loop to allow auto calibration to take place.