Which kind of information may be obtained by linking a computer to the optical socket of RadMan?
If you connect RadMan by optical link to the PC you will get ASCII values in Hex Format over the serial connector into your serial port of the PC. The ESM-TS software converts the Hex values into readable ASCII values. The software displays the measured value in % of standard, independently for the E and H field. The resolution is 0.625 % of standard from 0 % to 150 %. The rate is 25 measurements per second.