Which anesthesia gas machines require Vapor-Clean filters?
Published studies show that the newer Draeger anesthesia gas machines (Primus, Apollo, Fabius) and GE-Ohmeda (Aestiva) require 50-75 minutes of continuous gas flush at 10 L/min before residual vapor concentrations are less than the save level of 5 parts per million. It is likely that other new machines that contain plastic parts require similar preparation times.
Related Questions
- Will the QED-100 function properly if the capnometer/gas analyzer line is placed between the QED-100 and the anesthesia machine instead of between the patient and the QED-100?
- How soon after installation of the Vapor-Clean is the anesthesia gas machine free of trace anesthetic vapors?
- Do most dental procedures require Anesthesia?