What studies have been done on the petMAP?
A. This is a new product introduced in 2005. No outside direct arterial pressure (the gold standard) comparison studies have been done at this time. Dr. Maynard (Mike) Ramsey, M.D., Ph.D., the inventor of petMAP and a leader in blood pressure measurement device development, has extensive animal research experience. During the development of the petMAP product, he performed many animal BP comparisons in the laboratory of a well respected researcher at the University of Georgia Vet School. Based on those cat and dog experiments, Dr. Ramsey developed algorithms for both species and cuff site optimization to provide petMAP the best possible correlation with directly measured intra-arterial blood pressure (“the gold standard”). The result of these optimizations is that the BP measured with a petMAP will typically be 10-20% higher than those measured with other oscillometric devices and with Doppler devices. If the user wishes un-optimized readings, the BP can be displayed without optimizati