What should I do if there is no ECG trace visible on EPS320 Cardiac Stimulator screen although patient is connected?
• Select Alt-I, i.e. sense from catheter tip of paced electrode – this should always show an ECG if adequately placed in the heart. • If EPS320 is using External ECG inputs, verify connection between EP recorder and EPS320 Cardiac Stimulator and verify that EP Recorder has a sensing channel selected (e,g, highlight trace and press ‘s’ or similar). • Verify that ECG is indeed connected into the ECG inputs, and not the Sync Output on the rear of the SGU.