Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

What limits are placed on leakage and backscatter radiation?

0
Posted

What limits are placed on leakage and backscatter radiation?

0

The applicable performance standard for a diagnostic source assembly, under 21 CFR 1020.30(k), limits radiation leakage from the x-ray source assembly to an air kerma of 0.88 mGy (corresponding to an exposure of 100 mR) in one hour at a distance of one meter from the x-ray source when measured as specified in the standard.