What kind of uncertainties are provided with calibration reports issued by the Thermal & Mechanical Metrology program?
In the past, it was the practice of the Mass Standards program and its successor, the Thermal & Mechanical Metrology program, to specify uncertainties associated with values of variables as one standard deviation with an assumed normal or Gaussian distribution. This statement corresponds to a confidence level of approximately 68%. With the acceptance and publication of ISO Guide to the Expression of Uncertainty in Measurement (First Edition 1993) and with the growing international consensus on the provision of uncertainties with confidence levels of approximately 95%, this practice has changed accordingly. Unless otherwise explicitly stated, the uncertainties provided in calibration reports issued by the Thermal & Mechanical Metrology program shall correspond to a confidence level of approximately 95%. All such reports will bear a statement such as the following: Uncertainties specified in this report are expanded uncertainties representing a confidence level of approximately 95% obtai
Related Questions
- What kind of uncertainties are provided with calibration reports issued by the Thermal & Mechanical Metrology program?
- What kind of uncertainties are provided with calibration reports issued by the Mechanical Metrology Group?
- Are measurements made by the Thermal & Mechanical Metrology program traceable?