Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

What is the specificity and sensitivity of the VT/VF detection algorithm?

0
Posted

What is the specificity and sensitivity of the VT/VF detection algorithm?

0

The LifeVest meets the applicable portions of the AAMI standards for Automatic External Cardioverter Defibrillators “AECDs” (ANSI/AAMI DF2-1989 and AAMI DF39-1/93), which specify that the sensitivity for detecting VF should be greater than 90% and the sensitivity for detecting VT should be greater than 75%. It also meets the AAMI standard for specificity which states that the specificity for AECDs should be greater than 95% (without artifacts). ZOLL Lifecor bench testing demonstrated 100% detection sensitivity for VF and 95% detection sensitivity for treatable VT. Bench testing for specificity of the algorithm was 100%.