Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

What is the required upstream aerosol level for leakage testing?

0
Posted

What is the required upstream aerosol level for leakage testing?

0

Most current standards and recommended practices require a minimum of 10 micrograms per liter (ug/l) as an upstream aerosol challenge. While higher concentrations can be used, 10-ug/l is the minimum that may be used and still have a valid leakage certification test. All of ATI’s aerosol photometers were designed to operate at upstream challenge concentrations between 10 and 100 ug/l.