What is the optimum phased array coil design for cardiac and torso magnetic resonance?
To determine the optimum configuration of a phased array MR coil system for human cardiac applications, the sensitivity of 10 flexible array designs operating under ideal conditions was calculated at 13 points circling the myocardium of a model torso whose geometry was determined from healthy volunteers. The array geometries that were evaluated included continuous strips of 2, 4, 6, and 10 circular coils of diameter equal to half the torso thickness wrapped laterally around the torso, 2 pairs of coils located on the left side of the chest and back, clusters of 3 coils in 2 orientations, clusters of 4 and 6 coils, and a hybrid cross of 6 coils. The 4-, 6-, and 10-coil strip arrays out-performed the other designs for a given number of coils, yielding average theoretical sensitivity improvements of 45%, 53%, and 55% relative to a single flexible coil positioned at the point closest to the anterior myocardium, compared with about 30% for 4- and 6-coil clusters and the 2-pair geometry (P <