What is the minimum recommended frequency for performing ADR: Disinfection procedures?
a. An acid or vinegar descale are recommended to be performed following every Bicarbonate dialysis treatment. With version 3.33.3 software a brief integrated descale procedure occurs automatically following a bicarbonate dialysis treatment. A vinegar descale must be performed following the last treatment of the day. b. Perform at least one ADR: Heat or ADR: Chemical Disinfection procedure per day. c. If ADR: Heat is used, perform at least two ADR: Bleach Procedures per week to “deproteinize” the hydraulic pathway. d. Increase the frequency, change the method (Heat, Chemical), and/or change the disinfectant used during ADR: Chemical Procedures if the microbiological monitoring results do not meet applicable regulatory requirements or facility policies. Q. I am experiencing “! Flow Balance Error 144” and “W! Flow Balance Inaccurate 309” Alarms, especially later in the day, after several treatments have been performed. What can I do to prevent or reduce these alarms? A. These alarms are i