Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

What is the minimum recommended frequency for performing ADR: Disinfection procedures?

0
Posted

What is the minimum recommended frequency for performing ADR: Disinfection procedures?

0

a. An acid or vinegar descale are recommended to be performed following every Bicarbonate dialysis treatment. With version 3.33.3 software a brief integrated descale procedure occurs automatically following a bicarbonate dialysis treatment. A vinegar descale must be performed following the last treatment of the day. b. Perform at least one ADR: Heat or ADR: Chemical Disinfection procedure per day. c. If ADR: Heat is used, perform at least two ADR: Bleach Procedures per week to “deproteinize” the hydraulic pathway. d. Increase the frequency, change the method (Heat, Chemical), and/or change the disinfectant used during ADR: Chemical Procedures if the microbiological monitoring results do not meet applicable regulatory requirements or facility policies. Q. I am experiencing “! Flow Balance Error 144” and “W! Flow Balance Inaccurate 309” Alarms, especially later in the day, after several treatments have been performed. What can I do to prevent or reduce these alarms? A. These alarms are i