Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

What is the minimum/maximum reaction volume I should be using in the ABI PRISM® 7000 and ABI Prism® 7900HT?

0
Posted

What is the minimum/maximum reaction volume I should be using in the ABI PRISM® 7000 and ABI Prism® 7900HT?

0

For 96 wells plate, the minimum reaction volume is 25l for real time PCR and for Allelic Discrimination plate reads. For 384 wells plate, the minimum reaction volume is 10l for real time PCR and 5l for Allelic Discrimination plate reads. The maximum volume for 96 wells reaction is 50l and for 384 wells reaction is 20l.

Related Questions

What is your question?