What is the minimum interval between stimulation “pulses” using the DS5?
The DS5 will identify the end point on a stimulation pulse as the stimulus current value falling to within 400A for a minimum period of 200s or a sustained reversal of stimulus polarity (irrespective of the current amplitude). Therefore if you intend to carry out repetitive stimulation pulses of the same polarity, there needs to be a 200s gap between them. If the pulses are of alternating polarity, there is no requirement for an interval (e.g as in a sine wave).