Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

What is the minimum interval between stimulation “pulses” using the DS5?

0
10 Posted

What is the minimum interval between stimulation “pulses” using the DS5?

0
10

The DS5 will identify the end point on a stimulation pulse as the stimulus current value falling to within 400A for a minimum period of 200s or a sustained reversal of stimulus polarity (irrespective of the current amplitude). Therefore if you intend to carry out repetitive stimulation pulses of the same polarity, there needs to be a 200s gap between them. If the pulses are of alternating polarity, there is no requirement for an interval (e.g as in a sine wave).