What is the easiest way to check the dispensed volume of a syringe?
This is done by accurately weighing the amount of pure water that the syringe delivers (and recording the results a number of times). You will need an analytical, or semi-micro, balance in a temperature controlled environment, free from vibration. Draw up the water and dispense it exactly the same way a number of times. Wipe the outside of the needle so that no water adheres to it, taking care not to draw any water out of the needle itself.