Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

What is the dynamic range of the QuantiFluor™-P Fluorometer?

0
10 Posted

What is the dynamic range of the QuantiFluor™-P Fluorometer?

0
10

The dynamic range of the QuanitFluor™ Instruments is 4 logs, assay dependent. Dynamic range is often assay dependent because the assay background signal may limit the lower detection limit of the fluorophore and the fluorophore maximum intensity may limit the detection maximum of the fluorophore independent of the instrument.

Related Questions

What is your question?

*Sadly, we had to bring back ads too. Hopefully more targeted.