What is the drive volume of Bio-Logic stopped-flow?
The driving volume is freely and independently controllable form software and it depends on cuvette size. Mininum injection volume using 10 ml syringes is 50 μl. For high mixing ratio a small syringe can be installed without compromising dead time, and you can then inject 10-15 μl with one syringe. The Bio-Logic SFM allows samples to be saved by optimizing mixing sequence. With multiple syringe volumes to choose from and Bio-Logics completely independent syringe control, you are not limited to factory set 1:5 or 1:10. You can freely select the optimum ratio for your experiment ( if 1:7.6 is what you need, then just do it !