Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

What is the dead time between the data points during sequential store for the DL1540, DL1540L? ?

0
Posted

What is the dead time between the data points during sequential store for the DL1540, DL1540L? ?

0

The minimum dead time on the DL1540 and DL1540L is approximately 200 µs. However, this value varies depending on the T/Div setting, the number of channels, and the data length. Consider the minimum value to be the value obtained when the memory length is set to 1 KW and the T/Div is set to 1 µs/Div. For other settings, the dead time may even be a few milliseconds. Therefore, keep in mind that the value indicated above is a minimum value.