What is the correct power re-cycle process for the PCD 1000 Scanner?
On-going performance testing of the KODAK PCD 1000 Scanner has shown that there can be a significant improvement in the overall performance of the scanner if the following power re-cycling delay is observed. Important: Once you have turned the power switch to the “OFF” position, do not return the scanner power switch to the “ON” position until at least 100 seconds have passed. Following this 100 second delay in re-establishing power to the KODAK PCD 1000 Scanner will help enhance the overall life of the scanner. • What can I do about “Invalid Exposure” errors? Use the following sequence of corrections to fix Invalid Exposure errors (error code numbers 14039, 14288, 14289, 14290, and 14291). • Replace the lamp. • Clean the lens and light path. Note: the lens and light path include the following items: • The lens by the lamp – make certain it is not cloudy • The filter wheel • The light bar • Check that the Scan Lamp is seated correctly in the lamp socket. Occasionally, the pins have bee