Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

What is the correct calibration technique for the Model 935 MCA in Manual Mode?

0
Posted

What is the correct calibration technique for the Model 935 MCA in Manual Mode?

0

The correct procedure to calibrate the SAM is to let the auto calibrate feature raise the voltage until the PMT gain is in the linear region with proper course/fine gain to perform the required response with Cs137. According to the detector manufacturer‘s recommendation this should be between 600 – 900 volts (including the 4x4x16 inch Sodium Iodide detector). The SAM will automatically do this – manual adjustments are unnecessary.