What is the cause of the EE Error Indicator?
The most common error symbol that appears on the display panel is a EE with a down arrow and two numerals. The reason for the EE Error Indicator is generally caused because the cuff has been wrapped too loose. The cuff was under-inflated during the measurement. This error will also display if the cuff is worn and needs to be replaced. The EE symbol with the down arrow indicates the cuff is over-inflated. In most instances the cuff has been wrapped too tight or you are using the incorrect cuff size. Turn the monitor off. Read the instructions to correctly apply the cuff. Wait five to ten minutes than take another measurement. Error indicators vary by model number. Symbols that display on your model are printed in the instruction manual with the steps to correct the problem.
The most common error symbol that appears on the display panel is a EE with a down arrow and two numerals. The reason for the EE Error Indicator is generally caused because the cuff has been wrapped too loose. The cuff was under-inflated during the measurement. This error will also display if the cuff is worn and needs to be replaced. The EE symbol with a down arrow indicates the cuff is over-inflated. In most instances the cuff has been wrapped too tight or you are using the incorrect cuff size. Turn the monitor off. Read the instructions to correctly apply the cuff. Wait five to ten minutes then take another measurement. Error indicators vary by model number. Symbols that display on your model are printed in the instruction manual with the steps to correct the problem.