Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

What is the Calibration Period of the IBP Dialysis Meters?

0
Posted

What is the Calibration Period of the IBP Dialysis Meters?

0

The calibration period is one year. We recommend to calibrate the conductivity once per month with a conductivity reference solution. The monthly test recommendation is base on the experience that some technicians do not clean the electrode frequently. If biological material grows inside the cell it could interfere the accuracy of the conductivity measurement.