What does RenaiSys do during a preventative maintenance visit?
Below is a copy of our checklist: 1. Check with operator about backup, expectations, and known problems 2. If software maintenance is included, check on backup and install updates 3. Initialize system and run diagnostics program for instrument and its peripherals 4. Check optics (oculars, doves, zooms, POS/NEG, ortho/pseudo, and measuring marks) 5. Check illuminations 6. Check limit and end switches 7. Check MMA, PFKB, Table, Digitizer, Printer, and other connected peripherals 8. Repair any reported problem 9. Remove covers 10. Check magnets and moving cables 11. Check and if needed tighten screws and nuts (zoom, rheostats, …) 12. Clean and lubricate bearings and running surfaces 13. Clean optics, image carriers, and covers 14. Check and if needed adjust power supply and electronics 15. Install covers 16. Clean work area, monitor, keyboard, and other external components 17. Initialize system and run diagnostics program for instrument and its peripherals 18. Check calibration and if n