Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

What does Error Code 27 on the 8970B Noise Figure Analyzer indicate?

0
Posted

What does Error Code 27 on the 8970B Noise Figure Analyzer indicate?

0

The Agilent 8970B/71B/71C Noise Figure Measurement Operating Manual specifies the error code as: “Overflow while calibrating”. This error is generated when excessive loss exists between the noise source and the noise figure analyzer´s input during calibration. This condition could exist due to several conditions as outlined below: High insertion loss in the calibration path due to external high loss devices such as an: isolator, attenuator, cable, etc Noise source not attached to the analyzer’s input A failed noise source Improper switching of the noise source +28V dc control voltage Other service failures If excessive insertion loss is the culprit, remove the path loss from the calibration path, utilize Special Function 34.x, “Loss Compensation”, to account for the pre and/or post DUT insertion loss, and re-calibrate directly between the noise source and analyzer input. Another alternative is to implement Special Function 32.