What components of an old QuikChem system can be used for an upgrade to the QuikChem 8500 Series 2 FIA+?
XYZ (8000 and AE only) samplers can be refurbished as part of the upgrade. The same is true for pumps on either QuikChem IV, AE or 8000 systems. Chemistry manifolds are exchangeable from the AE and 8000 to the QuikChem 8500 platform, but not from the QuikChem IV. Please note, however, that AE users wishing to use their original manifolds will need to purchase new interference filters.