What can be used to dilute the StabilZyme Stabilizers?
We recommend using the stabilizers undiluted for maximum stabilization. However, in cases where you feel it may be necessary to dilute the products (e.g., 1:1) please use the following diluents: • StabilZyme HRP Stabilizer can be diluted in deionized water. • StabilZyme AP Stabilizer can be diluted in deionized water or 10mM MES buffer, pH 6.0. • StabilZyme SELECT Stabilizer can be diluted in deionized water or 0.1 M Tris or Glycine buffer. • StabilZyme NOBLE Stabilizer can be diluted in deionized water. NOTE: Phosphate buffers cause precipitation in the StabilZyme products when used at >10% total volume.