What are the volumes of Glen synthesis columns?
RESPONSE: TWIST (1.0m/0.2m/40nm) (20-0030-00) dimensions of inner chamber: diameter= 0.235 inches or 0.5969cm radius= 0.1175 inches or 0.2985cm (1 inch = 2.54cm) height= 0.320 inches or 0.8128cm Volume= πr2h 0.0139in3= 3.14 x 0.0138 x 0.320 or 0.227cm3= 3.14 x 0.0891 x 0.8128 (0.23ml) This calculated volume was checked by using a microliter syringe and found to be accurate. TWIST (10m or 15m) (20-0040-00) dimensions of inner chamber: diameter= 0.85-0.9cm (tapered) (use 0.875cm for diameter) radius= 0.4375cm height= 2cm Volume = πr2h 1.2cm3= 3.14 x 0.1914 x 2 (1.2ml) This calculated volume was checked by using a 5cc syringe and found to be accurate. The inner volume for the crimp-style 10/15m column (20-0020-05) is also 1.2ml. Perseptive/Expedite (0.2um) (20-0020-02) Volume was determined to be 40l or 0.04ml. (A microliter syringe was used to estimate the internal volume of this column.) Perseptive/Biosearch (1 .0um) (20-0020-01 ) Volume was determined to be 120l or 0.12ml. (A microlite