What are the time weightings in the Data Capture dialog?
When Jade 2 calculates the Lp values for storing in the data file, it uses a time weighting. The two time weightings provided are the standard ones of Fast and Slow (with a time constant of 125 milliseconds and 1 second respectively). Note that these time weightings are independent of any time weighting set on the meter. For most purposes, the Fast setting is preferred. However, certain statistics, such as Losha and Ldod, require the Slow setting.