Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

What are the settings for yeast lysis with the BigPrep Adapter?

adapter lysis settings yeast
0
Posted

What are the settings for yeast lysis with the BigPrep Adapter?

0

The FastPrepĀ®-24 instrument in combination with the BigPrep Adapter has been successfully used for the lysis of Pichia Pastoris, a yeast strain that has a big cell wall. After centrifugation of 1 liter culture (5x 10e8 cells/ml), the cell pellet is resuspended in 50 ml Lysis Buffer (containing protease inhibitors and PMSF). 25ml of this solution is added to 2 Big Lysing Matrix B tubes (containing silica beads) and samples are homogenized 4 times for 30 seconds at speed 6.0 m/s. Tubes were incubated on ice for 2 minutes between each run. 80% of yeast cells were lysed.