Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

What are the restrictions for hooking up to the process vacuum?

hooking process vacuum
0
10 Posted

What are the restrictions for hooking up to the process vacuum?

0
10

The process vacuum supplied to the labs (“house vac” or “process vac”) is only intended for dry systems. NEVER attach the process vac to a solution for degassing. Moisture and vapors that are drawn into the process vac will damage the facility pumps and shut down this support capability for all of BNC. Please be cautious when hooking up to this system and contact one of the BNC Managers if you need assistance setting up a vapor trap. Note: Any glassware or equipment that is used with the Process Vac must be shielded and/or explosion proof.