Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

What are the requirements for documenting manual integrations for initial and continuing calibration (Box #50 and #57, Sections 5.5.2.2.1 and 5.5.10)?

0
Posted

What are the requirements for documenting manual integrations for initial and continuing calibration (Box #50 and #57, Sections 5.5.2.2.1 and 5.5.10)?

0

EDQW Response: The data reviewer or user must be to be able to completely reconstruct chronological events leading to the final manually integrated chromatograms. Many software packages for audit trails do not allow full reconstruction and capture of “before” chromatograms. In those cases “before” chromatograms must be provided along with other information specified in DoD Box 50.