Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

What are the repeatability and reproducibility of the MS-200?

0
Posted

What are the repeatability and reproducibility of the MS-200?

0

Each MS-200 is required to pass a factory acceptance test specifying repeatability of better than 10% for 3 consecutive measurements of 1 ppm benzene in a nitrogen standard, with a background check between measurements. Typical achieved performance is better than 5%. Also, a linearity of response of better than 10% is required from mixtures of benzene in nitrogen over the range 200 to 1000ppb of benzene. The MS-200 is designed to operate in ambient temperatures up to 35°C. To maintain constant permeation properties of the membrane, the inlet is operated at approx 45°C to minimize the effects of ambient temperature fluctuations over the range 15-30°C. The temperature is controlled to within 0.1 degrees around the set value. At ambient temperatures of ~35°C the membrane temperature has to be set higher. The maximum temperature set point of the membrane in the present system is 90°C.