Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

What are the OSHA fume hood face velocity requirements?

0
Posted

What are the OSHA fume hood face velocity requirements?

0

OSHA does not have specific requirements for fume hood face velocity. The most widely accepted references on this subject ANSI/AIHA Z9.5 American National Standard for Laboratory Ventilation and the National Research Council’s Prudent Practices in the Laboratory recommend face velocities of 80 to 120 feet per minute (FPM).