What are the MVS performance specifications for accuracy and precision?
The MVS performance claims using the 96-well ARTEL Verification Plate for accuracy and precision are less than 3% and less than 1% CV, respectively. When using the 384-well ARTEL Verification Plate the performance specifications are 5% and 2% CV for accuracy and precision respectively. For a detailed uncertainty budget, contact Technical Support from 8:00 am to 5:00 pm Eastern Time at 888-406-3463 x109 or via e-mail.