What are the FDA guidelines for stent testing?
Accelerated in vitro testing for approximately 10 years of equivalent real time should be conducted. A statistically signifiant sample of stents expanded to their largest intended diameter and dynamically cycled using simulated vessel conditions should be tested. A complete description of the test protocol and sample preparation used in this study should be provided.