Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

What are the electrical lengths of the long transmission lines on my impedance standard substrate?

0
Posted

What are the electrical lengths of the long transmission lines on my impedance standard substrate?

0

If TRL calibration is used the delay of the long lines will need to be supplied to WinCal or directly to the VNA. These delays are readily determined from the drawing supplied with the ISS. Simply measure the line using a scale provided on the ISS for this purpose. Remember to allow for the probe contact overlap – about 50 um on each end. Each 130 um (0.130 mm) of length corresponds to approximately 1 ps, so delay may be calculated from T (ps) = Length (mm)/0.13 (best fit).

Related Questions

What is your question?

*Sadly, we had to bring back ads too. Hopefully more targeted.

Experts123