Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

We are servicing a 3000P analyzer and have found the sample inlet fitting at the rear of the instrument contains a small sintered filter that is blocked. What part is this?

0
Posted

We are servicing a 3000P analyzer and have found the sample inlet fitting at the rear of the instrument contains a small sintered filter that is blocked. What part is this?

0

That would be the flow restrictor. Replace it with P/N R2324 (install at inlet). The restrictors are also included with the restrictor Kit p/n A68729. Refer to Appendix page A-7 in the 3000PA instruction manual.