Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

There is only one analog output on the SBA-4 board and it is used to output CO2. How often (time interval) is the CO2 analyzer sampled?

0
Posted

There is only one analog output on the SBA-4 board and it is used to output CO2. How often (time interval) is the CO2 analyzer sampled?

0

A new CO2 value is calculated and output to the analog and digital outputs at 1.6 sec intervals. The internal sampling rate of the sensor is higher, but response time of the SBA-4 is typically limited by the time it takes for new gas sample to flow through and fill the sampling cell. The cell is only 2.5 cc and can be filled in 0.3 secs at a flow rate of 500 cc/min, but the tubing from the user environment to the cell can easily increase that time by an order of magnitude.