Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

The value of O2 being reported by my unit seems wrong. What might cause an erroneous reading?

0
0 Posted

The value of O2 being reported by my unit seems wrong. What might cause an erroneous reading?

0
0

Check the reference gas flowrate. The reference gas flowrate needs to be between 0.1 to 0.5 liters per minute. If it is outside of this, significant errors can result in the O2 reading. Also check that the sample flow rate is between 5 to 10 liters per minute.