Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

The standard analog repeat output on a series 40B display is 4…20mA, but my application requires 0…20mA. What can I do?

0
Posted

The standard analog repeat output on a series 40B display is 4…20mA, but my application requires 0…20mA. What can I do?

0

You have to enter the user calibration (U.C.) menu, set code 1, connect a milliammeter to the output, and when the message C.L. (minimum calibration) appears, set the read value to 0mA by means of the UP-DOWN keys.