Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

The Q-100, Q400, or Q-500 dosimeter displays DFx or DEFAULT x on the power up instead of the normal Inl or INITALIZING; Why?

0
Posted

The Q-100, Q400, or Q-500 dosimeter displays DFx or DEFAULT x on the power up instead of the normal Inl or INITALIZING; Why?

0
Anonymous

Nice share… Thanks for sharing

0

Depending on the default number, x, this is an indication that peak calibration constants have been lost, the serial number has been lost, and/or the setup information has been lost. The most likely cause for this is that the lithium battery inside the meter is dead. That lithium battery backs-up the above information when the 9Volt battery is removed. To correct the default condition the lithium battery should be replaced. Then the lost information must be restored. The end user cannot replace the lithium battery this is done by our service technicians.