Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

The data sheet suggests reading at 570 nm, but my instrument can only read at 550 nm?

0
Posted

The data sheet suggests reading at 570 nm, but my instrument can only read at 550 nm?

0

A. Please check the band width of the instrument. If it is 550+/- 20 you will still be in the wavelength range where you can estimate the amount of sample. In this case you should be fine as the equipments can measure at 550+20. In some other cases if you have to measure at 585 nm and the data sheet suggest 570 nm, you may be able to capture only 70-80% of the signal.